Adsorptive Apheresis Monitor Adamonitor SC
Adamonitor is the Adsorptive Apheresis Monitor used with Adacolumn (a granulocyte and monocyte/macrophage adsorption device, by JIMRO Co., Ltd.). Adamonitor is an adsorptive apheresis device exclusively for apheresis at a low flow rate. |
- Features
- Specifications
Features
- Capable of maintaining a stable flow rate even at a low flow rate.
- Safety mechanism activated when various sensors detect abnormalities.
Specifications
Specifications of Device | |
Rated voltage | 100-240 V AC |
Rated frequency | 50/60 Hz |
Power consumption | 200 VA |
Dimensions | 230(W) × 328(D) × 515(H) mm |
Weight | Approx.11kg |
★Blood Pump | |
Pump type | Two-roller pump |
Present flow rate range | Circulation stage: 10 to 50 mL / min Preparation stage: 10 to 150 mL / min |
Time setup range | 1 to 180 minutes |
Safety mechanism | Pump cover operation sensor Rotation Sensor |
★Syringe Pump | |
Present flow rate | 1 to 20 mL / h |
Safety mechanism | Syringe sensor Syringe movement sensor Syringe completion sensor Plunger set sensor Plunger overload sensor Feed screw and nut of opening and closing sensor |
Applicable syringe | TERUMO, NIPRO, JMS, TOP (20mL luer-locking type plastic syringe) |
★Sensor | |
●Normal Saline Solution Outage Sensor | |
Detection system | Ultrasonic system |
●Bubble Sensor | |
Detection system | Ultrasonic system |
Detection sensibility | Closed cell of 0.1 mL or over |
Safety mechanism | When the bubble sensor detects bubbles, the blood pump is immediately shut down, the bubble circuit is closed by the clamper, the abnormality is reported with warning buzzer, warning lamp, and monitor display. Almost no continuous air infusion measured (much lower than the value of 0.03mL/min). |
●Internal Pressure of Venous Circuit | |
Measurement display range | ‐13.3 to 53.3 kPa(-100 to 400 mmHg) |
Venous pressure warning setup range |
‐13.3 to 53.3 kPa(-100 to 400 mmHg) |
Safety mechanism | When the pressure in the dip chamber or the venous circuit exceeds the venous pressure warning setup range, the blood pump is stopped automatically, and the abnormality is reported with the warning buzzer, warning lamp. |
●Pillow Sensor (Negative Arterial Pressure Sensor) | |
Detection system | Mechanical switch |
Safety mechanism | When the abnormality (the pressure in the pillow becomes negative and the pillow shrinks due to blockage in blood vessel, etc) is detected by the negative arterial pressure sensor, the blood pump is immediately stopped, the blood circuit is closed by the clamper, and the abnormality is reported with the warning buzzer, warning lamp. |